Parameter assignment

9.1

Multipoint calibration

The transmitter provides an additional 20 point piecewise linear calibration table with user
selectable input value: volume flow, reynolds number or fluid sound velocity. The table can be
configured for either unidirectional or bidirectional calibration.
Note
Ascending order by reynolds number
The datapoints must be entered in ascending order by reynolds number.