Path Diagnostic Alarm Limits - Siemens FST020 Function Manual

Sitrans f series ultrasonic flowmeters ip65 nema 4x
Table of Contents

Advertisement

Related
Path
3.1.7.3

Path diagnostic alarm limits

3.1.7.3.1 Percent bursts accepted alarm limit
Description
Sets the desired limit for the percentage of accepted bursts. If measured percentage
accepted burst drops below this limit for any installed path, an alarm will be issued.
Setting
Default
Related
Path
3.1.7.3.2 Correlation factor alarm limit
Description
Sets the desired limit for the correlation factor. If the measured correlation factor drops below
this limit for any installed path, an alarm will be issued.
Setting
Default
Related
Path
3.1.7.3.3 RxGain alarm limit
Description
Sets the limit for the RxGain value. If the RxGain goes above this limit for any installed path,
an alarm will be issued.
Setting
Default
FST020 Function manual
Function Manual, 09/2017, A5E42949526-AA
Setup → Sensor settings → Path settings → Path 1 → Serial number
Min: 0%; Max: 100%
50%
Setup → Sensor settings → Path settings → Path diagnostic alarm limits → Percent bursts
accepted alarm limit
Min: 0.0; Max: 1.0
0.95
Setup → Sensor settings → Path settings → Path diagnostic alarm limits → Correlation factor
alarm limit
Min: 0.0 dB; Max: 100.0 dB
50.0 dB
Setup
3.1 Sensor settings
31