Calibration Point; Multipoint Calibration Table - Siemens FST020 Function Manual

Sitrans f series ultrasonic flowmeters ip65 nema 4x
Table of Contents

Advertisement

Setup
3.1 Sensor settings
3.1.9.3

Calibration point

Description
Calibration point that is to be transferred to the multipoint calibration table.
Setting
Default
Related
Path
3.1.9.4

Multipoint calibration table

3.1.9.4.1 Calibration points 1 - 20
Description
Calibration points 1 through 20 for unidirectional flow or calibration points 1 through 20 for
bidirectional positive flow. If 0.0 the calibration point is not used.
Access level: Expert user only.
Note
Only available for PID_ProtectCalibrationTable == 1 OR PID_EnableUserCalibrationTable ==
1
Setting
Default
Related
Path
3.1.9.4.2 Calibration value
Description
Calibration correction factor that is associated to Calibration points 1 through 20.
Access level: Expert user only.
Setting
Default
34
0.0 m
/s to 100000.0 m
/s
3
3
0.0 m
/s
3
Setup → Sensor settings → Calibration point
0.0 m
/s to 100000.0 m
/s
3
3
0.0 m
/s
3
Setup → Sensor settings → Multipoint calibration → Multipoint calibration table → Calibration
points 1 - 20
0.5 to 2.0
1.0
FST020 Function manual