3.
If Short + Long Press button is selected, assign desired
action to Short Press and Long Press.
a.
If actuator motion is assigned to parameter, Actuator
Operation Mode must be selected.
The options are:
Extend
Retract
Latched Extend
Latched Retract
Toggle
Latched Toggle
4. If Momentary button is selected, assign desired action
to button.
a. If actuator motion is assigned to parameter, Actuator
Operation Mode must be selected.

9.11 Installing / Setting up alternative inputs

There are different types of alternative inputs, that interface
differently to the system. Connect the components to the
particular inputs, see table below.
1605129-C
Modifying wheelchair configuration with PC tool
Component
Sip and Puff
Sip and Puff Head Array
Pediatric Compact Joystick
Head Array
Chin control
Compact Single Switch
Joystick
Four Switch Proximity Array
Remote Stop Switch
To install a Remote Stop Switch, just connect to the
Power Module using the interface cable and to the
4–way connector using the Bus cable, see wiring
diagram 3.8 Wiring diagrams, page 16. Programming
is not necessary.
Programming
1. Open Functions.
2. Add new profile or add user function to existing
profile. For more information about adding profiles and
functions, refer to 9.1 Profile and function actions, page
77.
3.
Set Function User Input to Input Module or to Third
Party, depending on component input.
Same procedure for drive, seating and
connectivity functions.
4. Open Chair Setup –> Modules –> IN 5xx or TPI,
depending on component input.
Input
Input Module IN5xx
Input Module IN5xx
Input Module IN5xx
TPI
TPI
TPI
TPI
Power Module
87