Invacare LiNX Service Manual page 89

Controls system for wheelchairs
Hide thumbs Also See for LiNX:
Table of Contents

Advertisement

Minimum limit for soft puff,
minimum limit for hard puff,
minimum limit for soft sip and
minimum limit for hard sip.
To calibrate the Sip and Puff module, you have to record
pressure measurements from a series of pressure samples
performed by the user and adjust the four pressure
thresholds accordingly.
1.
Open Connection in menu bar.
2. Select Sip and Puff Calibration.
3.
A Hard zone for puff input
B Threshold slider to adjust minimum limit for hard
puff
C Soft zone for puff input
D Threshold slider to adjust minimum limit for soft
puff
E Neutral zone for puff input
F Neutral zone for sip input
G Threshold slider to adjust minimum limit for soft
sip
H Soft zone for sip input
I Threshold slider to adjust minimum limit for hard
sip
J Hard zone for sip input
1605129-C
Modifying wheelchair configuration with PC tool
4. User performs hard puff, soft puff, hard sip and soft sip
in no specific order. For each sip and puff, a pressure
level K is recorded on calibration tool. Based on these
recorded levels, you determine best level at which to
set each pressure threshold.
5. Drag corresponding Threshold sliders.
For soft puff, drag threshold slider D to position
just below user's soft puff. Any subsequent pressure
below threshold slider D is in neutral zone E and
is ignored. Pressure above threshold slider D and
below threshold slider B is in soft zone C and
determined a soft puff.
For hard puff, drag threshold slider B to position
just below user's hard puff. Pressure below
threshold slider B is not considered as hard puff.
Pressure above threshold slider B is in hard zone
A and determined as hard puff.
For soft sip, drag threshold slider G to position just
above user's soft sip. Any subsequent pressure
above threshold slider G is in neutral zone F and
is ignored. Pressure below threshold slider G and
threshold slider I is in soft zone H and determined
as soft sip.
For hard sip, drag threshold slider I to position just
above user's hard sip. Pressure above threshold
slider I is not considered as hard sip. Pressure
below threshold slider I is in hard zone J and
determined as hard sip.
Sip and Puff Ramp Time
The Sip and Puff Ramp Time defines the amount of time
that a pressure input must remain in the soft zone before
it is registered as a soft sip or puff. Consequently, it also
provides:
Time for a user to cross from the neutral zone to the
hard zone and
a filter for any noise (unintentionally pressures) at the
zone boundaries.
The ramp time C begins at the point where the pressure
input D, E or F crosses into soft zone H — either from
neutral zone I, when pressure is increasing, or from hard
zone G, when pressure decreases. A pressure input D
that reaches the hard zone before ramp time expires, is
registered as a hard sip or puff. A pressure input E within
soft zone at the end of ramp time is registered as a soft sip
or puff. Pressure inputs F that fall away into neutral zone
before ramp time expires are not registered as an input.
89