Invacare LiNX Service Manual page 90

Controls system for wheelchairs
Hide thumbs Also See for LiNX:
Table of Contents

Advertisement

Invacare® LiNX
90
When setting Sip and Puff Ramp Time, consideration should
be given to users, who cannot produce a hard sip or puff
within the ramp time. If ramp time is set too low, a user
may not be able to reach the hard zone G in time and a soft
sip or puff is registered instead of the intended hard sip or
puff. For users who need more time to reach hard zone, the
ramp time C needs to be extended, so the pressure input
J is registered as a hard sip or puff.
1. Open Chair Setup –> Modules –> IN 5xx.
2. Adjust Sip and Puff Ramp Time.
1605129-C