Invacare® LiNX
1. Open Chair Setup –> Modules –> desired input to which
external stop switch is connected.
2.
Set Input Type to desired input, for example,
Momentary.
3.
Click on information icon next to Momentary.
Enable Stop driving and seating.
You do not need to assign a function to the
switch.
Configure latch drive parameters
The following parameters can be set individually for each
drive function.
82
1. Open Functions –> open desired drive function.
2.
Set type of latched driving with Latch Driving. For more
information about types of latched driving, refer to user
manual of remote.
3.
Set Latch Driving in Reverse parameter to On, to allow
latched driving in reverse.
4. Set maximum time that driving is latched without input
from user with Latch Drive Timeout parameter.

9.7 Adjust quadrant operation

This parameter sets the drive function's user input to operate
in 3–quadrant (3Q)or 4–quadrant (4Q) mode. Typically, the
4Q mode is chosen with joystick-based inputs or arrays, that
provide 4 quadrants, for example, a Sip and Puff array. The
3Q mode is chosen for user inputs that provide 3 quadrants,
for example, a head array.