7.7 Adjust quadrant operation

This parameter sets the drive function's user input to operate
in 3–quadrant (3Q)or 4–quadrant (4Q) mode. Typically, the
4Q mode is chosen with joystick-based inputs or arrays, that
provide 4 quadrants, for example, a Sip and Puff array. The
3Q mode is chosen for user inputs that provide 3 quadrants,
for example, a head array.
1. From Home screen open desired drive function.
The options are:
4Q Operation
3Q - Forward Only
3Q - Reverse Only
3Q - Manual Toggle
3Q - Automatic Toggle
4Q Operation
When 4Q Operation is selected, each quadrant is assigned a
unique direction from forward, reverse, left and right, this is
the default mode. To change these default directions, the
quadrants can be reassigned with a different direction by
setting the following parameters:
Forward Quadrant Operation
Reverse Quadrant Operation
Left Quadrant Operation
Right Quadrant Operation
It is not necessary that all quadrants are assigned unique
directions. If required, one or more quadrants can share the
same direction. For example, Forward Quadrant Operation
and Reverse Quadrant Operation can both be set to
Forward. For more information about adjusting the drive
direction, refer to 7.7.1 Adjust drive directions, page 65.
3Q - Forward Only
When 3Q - Forward Only is selected, the user input can
drive the wheelchair forward, left and right.
If required, each quadrant can be reassigned with a different
direction. Note that the Reverse Quadrant Operation
parameter is not applicable in 3Q - Forward Only mode.
3Q - Reverse Only
When 3Q - Reverse Only is selected, the user input can drive
the wheelchair reverse, left and right.
1605129-C
Modifying wheelchair configuration with iOS tool
If required, each quadrant can be reassigned with a different
direction. Note that the Reverse Quadrant Operation
parameter is not applicable in 3Q - Reverse Only mode.
3Q - Manual Toggle
When 3Q - Manual Toggle is selected, the user input can
drive the wheelchair forward, reverse, left and right. The
forward and reverse directions are controlled by the user
toggling an external switch.
If required, each quadrant can be reassigned with a different
direction. Note that the Reverse Quadrant Operation
parameter is not applicable in 3Q - Manual Toggle mode.
3Q - Automatic Toggle
When 3Q - Automatic Toggle is selected, the user input can
drive the wheelchair forward, reverse, left and right. The
forward and reverse directions are controlled by the user
toggling the forward quadrant.
If required, each quadrant input can be reassigned with
a different direction. Note that the Reverse Quadrant
Operation parameter is not applicable in 3Q - Automatic
Toggle mode.

7.7.1 Adjust drive directions

Forward Quadrant Operation
This parameter sets the direction that the wheelchair moves
when the forward quadrant of the user input is selected.
Options available are:
Forward
Reverse
Left
Right
Ignored
If Ignored is selected, the wheelchair does not move in any
direction when the input's forward quadrant is selected.
Reverse Quadrant Operation
This parameter sets the direction that the wheelchair moves
when the reverse quadrant of the user input is selected.
Options available are:
Forward
Reverse