Invacare® LiNX
1. Mount LiNX remote adapter B with screws A to remote
C.
The LiNX remote adapter can be turned 90° right
or left for more adjustment options.
2. Mount remote to Quad Link F with screw D, Nord-Lock
washer E and serrated lock washer G.
Tighten screw D with a maximum torque of 25
Nm.
Remote holder for attendant
The remote is fitted to the wheelchair using a bracket on the
push handle or the push bar.
•
3 mm Allen key
•
5 mm Allen key
•
8 mm wrench
30
1.
Mount remote A with support B and support drum C
to clamp D with screws F and washers E.
2.
Risk of damage to the remote
The maximum torque to tighten the screw
is 1 Nm.
– Do not exceed this rating as it may damage
the remote.
Fix support drum C in clamp D with screw G, washer
H and nut I.
3.
Attach clamp D to bracket J with screws L and
washers K.
4.
Attach bracket J to gripper clamp M with screws N
and nuts O.
3.9.4 Mounting the DLX-REM500
Swing away remote holder
•
3 mm Allen key
•
4 mm Allen key
•
Medium-strength thread-locking fluid, e.g.
Loctite 243
1605129-C