Performance Verification
14
18. Press the LIMITS softkey.
19. Press the SELECT softkey two times. Verify that SatSeconds SAT-S
alarm is selected.
20. Press the ADJUST UP button repeatedly and verify that the SatSeconds
alarm display cycles from OFF through 10, 25, 50, 100, OFF.
21. Press the POWER ON/OFF button to turn the monitor off.
22. Press the POWER ON/OFF button to turn the monitor back on.
23. Press the LIMITS softkey. Verify that the %SpO
display is boxed and indicates an alarm limit of "100."
24. Verify that the %SpO
alarm limit of "85."
25. Verify that the SatSeconds SAT-S alarm is set to OFF.
26. Verify that the pulse rate upper alarm limit display is boxed and indicates
an alarm limit of "170."
27. Verify that the pulse rate lower alarm limit display is boxed and indicates
an alarm limit of "40."
28. Press the POWER ON/OFF button to turn the monitor off.
lower alarm limit display is boxed and indicates an
2
upper alarm limit
2