Table 15: Electrostatic Immunity (IEC 60601-1-2)

The N-595 is suitable for use in the specified electromagnetic environment. The
customer and/or user of the N-595 should assure that it is used in an
electromagnetic environment as described below:
IEC 60601-1-
Immunity Test
1-2
Test Level
Portable and mobile RF communications equipment should be used no closer to
any part of the N-595, including cables, than the recommended separation distance
calculated from the equation appropriate for the frequency of the transmitter.
Radiated RF
3 V/m
IEC 6100-4-3
80 MHz
800 MHz
3 V/m
800 MHz
2.5 GHz
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Note: Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast, and TV broadcast cannot be predicted theoretically
with survey accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the N-595 is used
exceeds the applicable RF compliance level above, the N-595 should be
observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the
N-595.
Note: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
Interference may occur in the vicinity of equipment marked with the following
symbol:
Compliance
Electromagnetic
Level
Environment Guidance
Recommended Separation Distance
3 V/m
distance = 1.2√Power
3 V/m
distance = 2.3√Power
80 MHz to 2.5 GHz
3 V
distance = 1.2√Power
150 kHz to 80 MHz
Specifications
83