Patient Data and
Operating Status
Time
Patient Data
Operating Status
N-595
The Time column represents the N-595 real-time clock.
Patient data and the operating status of the unit are highlighted in the display
above. Parameter values, at the time of the printout, are displayed directly
beneath the heading for each parameter. In this example the %SpO
and the pulse rate (BPM) is 190 beats per minute. The asterisk (*) next to the
190 indicates that 190 beats per minute is outside of the alarm limits,
indicated in the top row, for pulse rate. If no data for a parameter is available,
three dashes (- - -) will be displayed in the printout.
Pulse Amplitude (PA) can range from 0 to 254. There are no alarm parameters
for this value. It can be used for trending information and is an indication of a
change in pulse volume, pulse strength, or circulation.
The Status column indicates alarm conditions and operating status of the
N-595. In this example the PH means Pulse High. The status codes are listed
in Table 19 on page 104. As many as 4 codes can be displayed at one time in
the Status column.
Data Port Interface Protocol
is 100,
2
103