Power-On Defaults
and Alarm Range
Limits
N-595
A 1-second beep sounds, indicating proper operation of the speaker,
and all indicators turn off except the AC Power/Battery Charging
indicator and the LCD screen.
The N-595 begins normal operation.
PLETH VIEW:
BLIP (MAGNIFIED) VIEW
Note:
When observing or changing alarm limits, a time-out is in effect
(approximately 10 seconds). If no action is taken within the time-out,
the monitor automatically returns to the monitoring display.
Note:
The descriptions that follow are based on the assumption that Pleth
view is the view that has been selected.
The steps for changing an alarm limit are the same if the view being used is
Blip (Magnified) view.
Note:
Power-on defaults will be the factory-set defaults or the defaults set
by your institution.
1. Turn on the N-595 by pressing the POWER ON/OFF button.
2. Press the LIMITS softkey. Verify that the monitor emits a single beep and
the pleth view is replaced with a display of the alarm limits. The upper
Performance Verification
11