Removal
60
10. Install the seven screws in the Power Supply and tighten.
11. Connect the fan harness to J1 on the Power Supply.
12. Install the PEM in the bottom case with the fuse drawer facing down. A
tab in the bottom case holds the PEM in place. Insert the bottom wing of
the PEM between the tab and the internal edge of the sidewall of the
bottom case. Push the PEM down and towards the outside of the monitor
until it clicks into place.
13. Complete the replacement procedure in paragraph Monitor Assembly on
page 54.
1. Complete the procedure in paragraphs Prior to Disassembly on page 52,
and Monitor Disassembly on page 53.
2. Disconnect the fan wire harness from J1 on the Power Supply PCB (see
Figure 13 on page 61).
3. Lift the cooling fan from the slots in the bottom case.