User Interface Pcb Removal/Replacement; Removal - Nellcor OXIMAX N-595 Service Manual

Hide thumbs Also See for OXIMAX N-595:
Table of Contents

Advertisement

Disassembly Guide
User Interface
PCB Removal/
Replacement

Removal

64
10. Complete the procedure in paragraph Monitor Assembly page 54.
1. Complete the procedures in paragraphs Prior to Disassembly page 52,
and Monitor Disassembly page 53.
2. Disconnect the CCFL harness (two white wires) from J5 of the User
Interface PCB. See Figure 14 on page 63.
3. Use a small blade screwdriver to pry the clip from either edge of J13, then
disconnect the Display PCB ribbon cable from the connector.
4. Disconnect the keypad ribbon cable from connector J17 on the User
Interface PCB (Figure 14 on page 63). Lift up on the ribbon cable's outer
shell until it clicks, then remove the cable from the connector.
5. Disconnect the speaker cable from J12 on the User Interface PCB.
6. Remove the five screws in the User Interface PCB (Figure 15 on
page 65).
7. Remove the User Interface PCB from the top case.