Performance Tests and Calibration

Calibration Process

N O TE
128
Now you can use 1234 as the security code. If you want to
enter a new security code, see
Calibration Security Code from the Front Panel"
page 126. Make sure you record the new security code.
To complete a full instrument calibration use following
general procedures:
1 Read
"Test Considerations"
2 Perform the verification tests to characterize the
instrument (incoming data).
3 Unsecure the instrument for calibration (see
Security"
on page 124).
4 Perform the adjustment procedures (see
Consideration"
on page 130).
5 Secure the instrument against calibration.
6 Note the new security code and calibration count in the
instrument's maintenance records.
Make sure to quit the Adjustment Mode before turning off the instrument.
Agilent U1251B/U1252B User's and Service Guide
"To Change the Instrument
on
on page 114.
"Calibration
"Adjustments