Calibration from Front Panel

Calibration process

N O T E
U1253B User's and Service Guide
The following general procedure is the recommended method
to complete a full instrument calibration.
1 Read and implement
2 Perform the verification tests (refer to
page 145) to characterize the instrument.
3 Perform the calibration procedures (refer to
procedures"
on page 164; read also
Considerations"
on page 158).
4 Secure the instrument after calibration.
5 Take note of the new security code (if it has been
changed) and the calibration count in the instrument
maintenance records.
Make sure to quit the adjustment mode before switching off the
instrument.
Performance Tests and Calibration
"Test Considerations"
Table 6- 2
"Adjustment
6
on page 142.
on
"Calibration
163