Calibration Process

The following general procedure is the recommended method to complete
a full instrument calibration.
1 Read
2 Perform the verification tests to characterize the instrument (incoming
3 Unsecure the instrument for calibration (see
4 Perform the adjustment procedures (see
5 Secure the instrument against calibration.
6 Note the new security code and calibration count in the instrument's
Make sure to quit the Adjustment Mode then turn off the instrument.
N O T E
34405A User's and Service Guide
"Test Considerations"
data).
page 82). Once unsecured, the instrument will be in Adjustment Mode
as indicated by the illuminated CAL annunciator.
maintenance records.
Performance Tests and Calibration
on page 67.
"Calibration Security"
"Adjustments"
4
on
on page 88).
85