Agilent MOI for MHL Cables Compliance Tests
16. Click Close.
17. Select
Trace
8.
18. Open
Parameters
19. Click Tcc22.
6.6. CBUS Cable Delay
1. Select
Trace
5.
2. Open
Parameters
3. Select "Time Domain" and "Single-Ended' for Measure.
4. Select Format to "Volt"
5. Click T21.
6. Click the box below the left knob under Vertical. Set vertical scale to "100 mV/div" in
a dialog box.
7. Click the box below the right knob under Vertical. Set vertical center to "100 mV" in a
dialog box.
The following procedures are manual setup for frequency domain measurements.
6.7. Common Parameters Setup for Frequency-domain Measurements
1. Press
Channel Next
2. Press
Sweep Setup
> Set start value to "900 kHz".
3. Press
Start
> Set stop value to "5.1 GHz".
4. Press
Stop
5. Press
Analysis
> Fixture Simulator and turn it ON.
6. Press
Analysis
> Fixture Simulator > Topology > Device > Bal-Bal
7. Press
Analysis
> Fixture Simulator > Topology > Port1 (bal) > 1-2
tab.
tab.
to select Channel 2.
> Points and set to "1,601".
32