Failing In Auto F.d.c. (Automatic Density Calibration) - FujiFilm Drypix 5000 Operation Manual

Table of Contents

Advertisement

Troubleshooting

5.6 Failing in Auto F.D.C. (Automatic Density Calibration)

If fails in automatic density calibration processing for any reason, an error window like that
shown below appears.
[Fig. 5.44]
Press the
If the same failure recurs after Auto F.D.C. processing is performed again, a film will be
printed by using a density table before the failure has occurred. If the density is normal,
continuous use is possible. If the error occurs again, report it to the nearest service center.
If fails in automatic density calibration processing after a film pack has been supplied, an
error window like that shown below will open.
[Fig. 5.45]
Press the
automatic density calibration processing.
5-34
button to perform automatic density calibration processing again.
button to chose it, and then Use Utility operation mode the perform
006-240-50 2006.10