Maximum Density; Power Supply Conditions; Environmental Conditions - FujiFilm Drypix 5000 Operation Manual

Table of Contents

Advertisement

Specifications

7 Maximum Density

For the DI-HL and DI-HLc, select one of 2.64, 3.0, 3.3 and 3.6.
For the DI-ML, select 3.6 or 4.0.
* Note that 3.6 and 4.0 are for mammography use only.

8 Power Supply Conditions

Input voltage
Phase
Frequency
Rated current

9 Environmental Conditions

(1) Operating conditions
Temperature : 15˚C (40%) – 30˚C (70%)
humidity
Atmospheric pressure : 700-1060hPa
Maximum heating value : 2100KJ
(2) Non-operating conditions (not including film)
Temperature
Humidity
Atmospheric pressure : 500-1060hPa
006-240-40 2006.09
AC100/110/120V ±10%
Single
50-60Hz
12A
: 15%RH (30˚C) – 70%RH (30˚C) (no dew condensation)
Humidity
70%
40%
15%
: 0 – 45˚C
: 10 – 90% RH (no dew condensation)
Temperature
1
2
3
4
5
6
7
7-3