Axiovert 200
LCD display (3-13/17)
Two-line display with 16 characters each
attached to the carrier for transmitted-light
illumination / condenser.
Display
of
objective
magnification (3-6/2) and contrasting technique
(3-6/3) in the upper line.
FL display (3-6/4) on the right of the top line if
the fluorescence shutter is opened.
Display of lamp brightness in a line scale and
display of the relevant voltage (3-6/5) in the
lower line. One interval corresponds to 0.4 V.
The LCD background illumination can be
switched on or off as required. Activation /
deactivation is performed by pressing the FL
key for approx. 1 second.
Brightness adjustment of the LCD display is by
pressing/holding the
.
s w i t c h
The
LCD
display
configuration of the microscope (Set mode).
Instead of the line scale for the lamp brightness, the following can be displayed in the lower line (3-6/6)
of the LCD display:
HAL off when the halogen illuminator is switched off.
3200K when the lamp voltage has been set to the color temperature (10.5 V).
Description of the swung-in filter set on the reflector turret (e.g. DAPI/FITC), provided that the
fluorescence shutter has been opened. If no filter set is contained in the reflector turret, the reflector
turret position (e.g. Refl. 2) is displayed instead.
The currently set position appears in the LCD display for 4 seconds during performance of the following
key functions: Both lines of the display are used.
1x Optovar: display of the Optovar magnification factor (1.6x) and the overall magnification (Total
Mag.) in the intermediate image
LEFT/RIGHT and Slider VIS/Frontport/Baseport: display of the actual intensities at Sideport,
Frontport, Baseport or VIS, depending on the selected positions of both control elements, e.g. SP:
Right 20% (upper line of the LCD display) and Binocular 80% (lower line).
FOCUS - Load: Load Position appears in the lower line (display blinks every second).
ZERO: display of the current focus position in mm (e.g. Zdis: +0.0µm)
If a motorized condenser is connected, the currently set condenser turret position appears in the
lower line (e.g. Cond. DICIII). If the aperture is changed by pressing the key on the condenser, the
value of the stop opening is displayed (e.g. Cond. NA: 0.55). When a phase stop position is set, the
aperture diaphragm is always entirely open.
Display function during the instrument configuration (SET mode).
B 40-080 e 03/01
OPERATION
Axiovert 200 M
position
(3-6/1),
key + the
F L
t o g g l e
is
also
used
for
the
Fig. 3-14
LCD-Display
Carl Zeiss
3-19