Contents Functions at a glance – VISU 200 BrightFlex™ surgical microscope – Key to symbols Safety – Directives and standards – Notes on installation and use – When using a fundus imaging system (e.g. BIOM II) – Phototoxic retinal injury in eye surgery –...
Page 4
Procedure What to do in an emergency – Failure of zoom system – Failure of X-Y coupling – Failure of focusing system Maintenance / Further information – Trouble-shooting table – Magnifications / Fields of view – Care of the unit –...
Functions at a glance Functions at a glance VISU 200 BrightFlex™ surgical microscope Key to symbols G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Functions at a glance VISU 200 BrightFlex™ surgical microscope Resetting the X-Y coupling, focus and zoom to their initial positions Page 30 Adjusting the tilt of the surgical microscope Page 50 Setting the interpupillary distance Page 40 Adjusting the eyecups Page 40 Setting your prescription Page 40...
The hand indicates hints on the use of the instrument or other tips for the user. ® OPMI ® OPMI is a registered trademark of Carl Zeiss. G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Page 8
Functions at a glance G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Safety Safety Directives and standards Notes on installation and use When using a fundus imaging system (e.g. BIOM II) Phototoxic retinal injury in eye surgery Warning labels and notes G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
– UL – CSA In accordance with Directive 93/42/EEC, the complete quality manage- ment system of the company Carl Zeiss has been certified by the DQS Deutsche Gesellschaft zur Zertifizierung von Managementsystemen mbH, a notified body, under registration number 250758 MR2.
Use this instrument only for the applications described. • Only use the instrument with the accessories supplied. Should you wish to use other accessory equipment, make sure that Carl Zeiss or the equipment manufacturer has certified that its use will not impair the safety of instrument.
Page 12
Safety • Only personnel who have undergone training and instruction are al- lowed to use this instrument. It is the responsibility of the customer or institution operating the equipment to train and instruct all staff using the equipment. • Keep the user's manuals where they are easily accessible at all times for the persons operating the instrument.
Page 13
Safety • Pay special attention to warning symbols on the instrument (triangular warning signs with exclamation marks), labels and any parts such as screws or surfaces painted red. For every use of the instrument • Avoid looking directly into the light source, e.g. into the microscope ob- jective lens or a light guide.
Safety When using a fundus imaging system (e.g. BIOM II) When using a fundus imaging system (e.g. BIOM II from the company Oc- ulus) which is usually installed between the surgical microscope and the patient, make sure that the patient is neither put at risk nor injured by the motorized focusing system or the movement of the stand arm.
Page 15
Safety In the following, comments on these aspects are given and a description of how Carl Zeiss, as a manufacturer, makes allowance for them in its in- struments. Illumination characteristics (spectral composition) Studies on exposure of the eye to light of varying spectral composition date back to the early 1950s.
Page 16
Also, the surrounding light sources should not cause addi- tional strain to the patient's eye. Carl Zeiss has provided an answer to this problem in the form of a swing- in retinal protection device for insertion into the beam path of the surgical microscope.
Page 17
Such factors would be expected to significantly extend the time at which photoretinitis might occur. In conclusion Carl Zeiss recommends: – Use of the GG 475 eye protection filter. – Reduction of the illumination of the surgical area to the extent required for the patient's safety and for microscopic visualization.
Safety Warning labels and notes Caution: Observe all warning labels and notes! If any label is missing on your instrument or has become illegible, please contact us or one of our authorized representatives. We will supply the missing labels. G-30-1439-en OPMI®...
Description Description VISU 200 BrightFlex™ Surgical Microscope Intended use Description of the modules Illumination system Controls, displays, connections Hand grips (option) Tube and eyepieces for main microscope Tube and eyepieces for assistant's microscope G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21.
Description VISU 200 BrightFlex™ Surgical Microscope Intended use The VISU 200 BrightFlex™ surgical microscope has been designed for surgical procedures in the field of ophthalmology, i.e. the microscope meets the special requirements of this discipline. Description of the modules The VISU surgical microscope is comprised of the following modules: 1 X-Y coupling The X-Y coupling allows fine, motorized positioning of the surgical mi- croscope in a horizontal plane.
Page 21
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Page 22
Description 2 Support arm for the surgical microscope The support arm incorporates a tilt device. This allows the viewing di- rection of the surgical microscope to be adapted to the requirements of the surgical field. Using the knob for fine tilt, you can position the surgical microscope in a range from +180°...
Page 23
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Page 24
Description To protect the patient's eye against photo-retinitis, a retinal protection de- vice is provided. This device can be swung into the illumination beam path, if no red reflex is required. At the light source integrated in the stand, a GG 475 eye protection filter can be swung into the illumination beam path.
Page 25
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Page 26
Description 4 0° assistant's microscope The assistant's microscope is an integral part of the VISU 200 Bright- Flex™ surgical microscope, i.e. it cannot be separated from the main microscope. The assistant sees the same image as the main surgeon. The system's excellent image quality is also available to the assistant. The assistant's microscope has two working positions.
Page 27
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Description Illumination system 6° illumination You can continuously fade out the 6° illumination. This results in markedly reduced reflection from the cornea, while providing nevertheless a high- contrast, informative microscope image. For cataract surgery, we recommend opening the 6° illumination system approx.
Page 29
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Description Controls, displays, connections 1 Instrument connector 2 Microscope shaft 3 X-Y coupling 4 Actuator button – recenters the X-Y coupling. – resets the focus to its initial position in the focusing range, and – the zoom system will set a preselected magnification factor if the XYZ-RES function has been activated (possible with S8 suspen- sion systems only).
Page 31
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Page 32
Description 9 Dust cover 10 Release buttons for the magnetic brakes of the stand Only in connection with suspension systems with magnetric brakes. A second button with the same function is located on the opposite side in the knob for the additional red reflex illumination. –...
Page 33
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Page 34
Description 14 Manual adjustment of zoom system 15 Knob for adjusting the red reflex illumination The angle of illumination can be switched between + 2° and - 2°. 16 Release button for magnetic brakes (A second button with the same function is located in maneuvering grip (10).
Page 35
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Page 36
Description 22 Light stop knob Retro-illumination contrast stop. This stop reduces the straylight reflected from the sclera. Diameter approx. 16 mm (with objective lens f = 200 mm): Free passage. Diameter over approx. 16 mm: Partially transmitting periphery. Free passage. The field of view is fully illuminated. Retinal protection device.
Page 37
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Description Hand grips (option) In addition to the maneuvering grips provided, larger hand grips can be mounted. The mounting procedure is described in the instructions en- closed. This does not affect the functioning of the surgical microscope. Only the operation of the surgical microscope is slightly changed. 1 Maneuvering grip with integrated release button By turning one of the two maneuvering grips you will activate the button for magnetic brakes of the suspension system.
Page 39
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Description Tube and eyepieces for main microscope 1 180° tiltable tube 2 PD adjustment knob The correct interpupillary distance has been set when both eyepiece images merge into one. 3 Push-in widefield eyepiece Note: When you remove the eyepiece from the tube, please note that this eyepiece is fitted with a magnetic coupling.
Page 41
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Description Tube and eyepieces for assistant's microscope 7 Push-in widefield eyepiece Note: When you remove the eyepiece from the tube, please note that this eyepiece is fitted with a magnetic coupling. When mounted, the eye- pieces display a very weak magnetic field. For this reason, the usual rules for the handling of magnets must only be observed with eye- pieces which have not been mounted on the microscope: •...
Page 43
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Page 44
Description G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Preparations for use Preparations for use Mounting the binocular tubes, eyepieces and the objective lens G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Preparations for use Mounting the binocular tubes, eyepieces and the objective lens Note: Always use the assistant's microscope with the appropriate binocular as- sistant's tube (shown). If you use a different tube, you will obtain an in- verted image. • Loosen securing screw (5) by a few turns.
Page 47
Preparations for use Caution: Before using and after re-equipping the unit, make sure that the two bin- ocular tubes (2 and 10) are securely locked in position. Make sure that – securing screws (5 and 7) and – objective lens (6) have been tightened firmly.
Page 48
Preparations for use G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Operation Operation Preparations for use Adjusting the tilt angle Setting the microscope tilt to angles greater than 15° Adjusting the surgical microscope Checklist When using a fundus imaging system (e.g. BIOM II) Procedure What to do in an emergency Failure of zoom system Failure of X-Y coupling Failure of focusing system G-30-1439-en...
Operation Preparations for use Adjusting the tilt angle Using knob (1), you can position the surgical microscope in a range from +180° to -180° (+ in the direction of the surgeon and - in the opposite di- rection). The +90° setting is ideal for surgery on patients in a seated po- sition or lying on their side.
Page 51
Operation G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Operation Setting the microscope tilt to angles greater than 15° • Using locking screw (2), secure the assistant's microscope in the working position set (on the left or right of the main microscope). • Check that the assistant's microscope is firmly seated. •...
Operation Adjusting the surgical microscope Bring the surgical microscope into starting position within the focusing range. Set the minimum magnification on the surgical micro- scope. Bring the surgical microscope into working position. Adjust the interpupillary distance on the binocular tube. Set the diopter scales on the eyepieces.
Operation Checklist Always check the following points before surgery (without the patient!): VISU 200 BrightFlex™ surgical microscope – Check that the correct rated voltage has been set for the suspension system. – Check that all cables have been connected. – Check that the light guide has been connected. •...
Page 55
Operation – Check that image quality is the same throughout the entire magnifica- tion range. Accessories – Using the relevant user's manuals, check that the other equipment (il- lumination system, video system, etc.) is working properly. Warning! If a function fails, you must not use this instrument for safety reasons. Cor- rect the fault (see the "Troubleshooting table“) or contact our service dept.
Operation When using a fundus imaging system (e.g. BIOM II) When using a fundus imaging system (e.g. BIOM II from the company Oc- ulus) which is usually installed between the surgical microscope and the patient, make sure that the patient is neither put at risk nor injured by the motorized focusing system or the movement of the stand arm.
Operation Procedure • Turn on the power switch of the suspension system. • Adjust the brightness of the surgical field illumination as required on the stand. • Check the unit using the checklist. • Move the surgical microscope over the surgical field and bring it into a convenient position within the working distance.
Operation Caution! – Avoid looking directly into the light source, e.g. into the microscope ob- jective lens or into a fiber light guide! – When selecting the brightness level for the patient's eye, always take care to keep the strain on the patient's eye to a minimum. –...
Operation What to do in an emergency Failure of zoom system • Manually adjust the magnification using zoom control (1) on the micro- scope (if necessary, use a tool, e.g. screwdriver, coin, etc.). G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21.
Operation Failure of X-Y coupling • If the X-Y coupling fails, you can manually position the surgical micro- scope utilizing the possibilities of the suspension system you are us- ing. Failure of focusing system • Disconnect the cable between the surgical microscope and the sus- pension system (illumination continues to be on.) •...
Maintenance / Further information Maintenance / Further information Trouble-shooting table Magnifications / Fields of view Care of the unit Sterilization Ordering data Spare parts Accessories Adapter cables G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Maintenance / Further information Trouble-shooting table • This instrument is a high-grade technological product. To ensure opti- mum performance and safe working order of the instrument, its safety must be checked once every 12 months. We recommend having this check performed by our service representative as part of regular main- tenance work.
Page 63
Maintenance / Further information Problem Possible cause Insufficient surgical field illu- Brightness level set too low. Adjust brightness on sus- mination. pension system or using foot control panel. Defective S light guide (no Contact service dept. homogeneous illumination). The light guide must prob- ably be changed.
Maintenance / Further information Magnifications / Fields of view Using the magnification factor γ of the zoom system, you can calculate the total magnification of the surgical microscope according to the following formula: tube ---------- - γ M ⋅ ⋅ where: is the focal length of the binocular tube tube...
To protect the eyepiece optics from fogging, we recommend using an anti-fogging agent. Note: Anti-fogging agents provided by eyecare professionals for use with eye- glass lenses are also suitable for Zeiss eyepieces. • Please observe the instructions for use supplied with each anti-fog- ging agent.
Maintenance / Further information Sterilization The asepsis sets available from Carl Zeiss contain rubber caps, sleeves and handgrips which can be sterilized in autoclaves. We recommend the following program for sterilization: Sterilization temperature: 134° C Sterilization time: 10 minutes Sterile single-use drapes are available to cover the instrument.
Maintenance / Further information Ordering data VISU 200 BrightFlex™ Surgical Microscope Description Cat.No. VISU 200 BrightFlex™ surgical microscope including 1179-101 X-Y coupling and assistant's microscope 180° tiltable tube 30 37 91-0000 10x eyepiece for assistant's microscope (2x) 30 55 42-0000 12.5x/18 B eyepiece for main microscope (2x) 30 55 43-0000 Objective lens, f = 175 mm...
Please observe the following: Only operate the instrument with the accessories included in the delivery package. If you want to use other accessories, make sure that Carl Zeiss or the manufacturer of the accessories has proved and confirmed that these accessories meet the respective technical safety standards and can be used without risk.
Page 69
Maintenance / Further information VISU 200 BrightFlex™ Surgical Microscope Description Cat. No. Objective lens, f = 200 mm 30 26 52-9903 Optical wedge, 30° 1006-145 Sleeve (for insertion of eyepiece (4)) 30 55 42-0107 10x eyepiece with magnetic coupling 30 55 42-0000 OPMI VISU hand grips 1049- 171 Asepsis for OPMI VISU hand grips...
Maintenance / Further information Adapter cables for the electrical connection of the surgical microscope to different types of suspension system. Cat. No. EMD II Electromechanical Ceiling Mount 30 26 02- 9053 S3 Ceiling Mount 30 26 02- 9053 S4 Ceiling Mount 30 26 02- 9054 S23 Ceiling Mount 30 26 02- 9053...
Technical data Technical data Technical data Ambient requirements G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Page 72
Technical data Technical data VISU 200 BrightFlex™ Surgical Microscope Optical data Magnification Main microscope: 4.9x - 29x Motorized zoom system with apochromatic op- tics, 1:6 zoom ratio, magnification factor γ = 0.4x - 2.4x. Focusing Motorized, focusing range 50 mm At the press of a button, the focusing drive returns to its initial position.
Page 73
Technical data Swing-in stops Patented retinal protection device, Pat. No. DE 33 39 17 2 C2 Patented field stop for reduced glare, Pat. No. G 91 03 43 3.7 Mechanical data Front-to-back tilt of With self-locking gear drive, manually adjustable microscope using a knob.
Technical data Ambient requirements For operation Temperature +10 °C...+40 °C Rel. humidity 30%...75% Air pressure 700 hPa...1,060 hPa For transportation and Temperature - 40 °C...+70 °C storage Rel. humidity 10%...100% (without condensation) Air pressure 500 hPa...1,060 hPa The unit meets the essential requirements stipulated in Annex I to the 93/42/EEC Directive governing medical de- vices.
Index Index Symbols +90° position Numerics 2° illumination Accessories Accident prevention regulations Actuator button Adapter cables Adjusting lever for 6° illumination Adjusting the microscope tilt to angles greater than 15° Adjusting the surgical microscope Adjusting the tilt angle Ambient requirements Angle of illumination Anti-fogging agent Arrows indicating the focusing range...
Page 76
Index Eyepiece optics, fog-free Failure of a function Failure of zoom system Fields of view Focus of the light source Focusing knob Focusing system failure Fogging of optical surfaces 14, 56 Fundus imaging system GG 475 eye protection filter GG 475 protection filter Hand grips (option) Illumination characteristics Installation, notes...
Page 77
Index Notes on safety Notes on use Operation Operation, requirements Optical surfaces, fogging Ordering data PD adjustment knob Phototoxic retinal injury in eye surgery Potential equalization Preparations for use Procedure Protective earth contact Red reflex illumination Re-equipping the instrument 34, 38 Release button for magnetic brakes Requirements for operation 14, 56...
Page 78
Index 20, 30 X-Y coupling X-Y coupling failure G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...
Page 79
G-30-1439-en OPMI® VISU 200 BrightFlex™ Surgical Microscope Issue 3.0 Printed on 21. 11. 2002...