Specifications
B-4
Power Outlets
A separate, dedicated power line with a 208 V, 15 amp dedicated receptacle is
required for the printer.
The outlets should meet the following standards:
Item
amperage
wire size
phase
receptacle
impedance between ground and
neutral wires
voltage drop across the circuit
breaker
distance between receptacle and
equipment
Line Frequency
In most cases, the electric utility company is responsible for maintaining the
proper frequency of the alternating current in its distribution network. If the
building is supplied with power by its own motor-generator, make certain that the
governor that controls the frequency is capable of delivering power within the
required operating frequency range.
Printer Standard
15 amp
14 gauge or larger
1 (single)
NEMA5—N6/15
less than 2 ohms
less than 0.1 volts
1.5 m (5.0 ft.)
or less
September 2000