Otto Bock Genium 3B1-3 Instructions For Use Manual page 47

Hide thumbs Also See for Genium 3B1-3:
Table of Contents

Advertisement

Beep signal
10x long
30x long
Error while charging the product
LED on
Status
power
LED on
supply
battery
charger
Vibration signal
10x long
1x long, 1x short
repeated every
3 seconds
Continuous
Error
Country-specific plug adapter not
fully engaged on power supply
Non-functional socket
Defective power supply
No connection between battery
charger and power supply
Defective battery charger
Event
Charge level 5%
After the beep and
vibration signals, the
product switches to
empty battery mode
and then switches
off.
Severe error/indic­
ation of safety
mode activation
(see page 39)
e.g. one or more
sensors are not oper­
ational
Total failure
Electronic control no
longer possible.
Safety mode active or
undetermined valve
state. Undetermined
product behaviour.
Resolution
Check whether the country-specif­
ic plug adapter is fully engaged on
the power supply.
Check socket with another electric
appliance.
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
Check whether the charging cable
plug is fully engaged on the battery
charger.
The battery charger and power
supply must be inspected by an
authorised Ottobock Service
Centre.
Required action
Charge the battery.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger.
If the error persists, use
of the product is prohib­
ited. The product must be
inspected by an O&P
professional immediately.
Attempt to reset this error
by connecting/discon­
necting the battery char­
ger.
If the error persists, use
of the product is prohib­
ited. The product must be
inspected by an O&P
professional immediately.
47