Be aware that resistance in the flexion and extension direction can change in case of warn­
ing and error signals.
Particular caution is required when carrying children down the stairs.
CAUTION
Overheating of the hydraulic unit due to uninterrupted, increased activity (e.g. extended
walking downhill)
>
Falling due to unexpected behaviour of the product because of switching into overheating
mode.
>
Burns due to touching overheated components.
Be sure to pay attention when pulsating vibration signals start. They indicate the risk of over­
heating.
As soon as these pulsating vibration signals begin, you must reduce your level of activity so
the hydraulic unit can cool down.
Full activity may be resumed after the pulsating vibration signals stop.
If the activity level is not reduced in spite of the pulsating vibration signals, this could lead to
the hydraulic element overheating and, in extreme cases, cause damage to the product. In
this case, the product should be inspected for damage by an O&P professional. If neces­
sary, they will forward the product to an authorised Ottobock Service Center.
CAUTION
Overloading due to unusual activities
>
Falling due to unexpected product behaviour as the result of a malfunction.
>
Falling due to breakage of load-bearing components.
>
Skin irritation due to defects on the hydraulic unit with leakage of liquid.
The product was developed for everyday use and must not be used for unusual activities.
These unusual activities include, for example, extreme sports (free climbing, paragliding,
etc.).
Careful handling of the product and its components not only increases their service life but,
above all, ensures your personal safety!
If the product and its components have been subjected to extreme loads (e.g. due to a fall,
etc.), then the product must be inspected for damage by an O&P professional immediately.
If necessary, they will forward the product to an authorised Ottobock Service Center.
CAUTION
Improper mode switching
Falling due to unexpected behaviour of the product because of changed damping behaviour.
Ensure that you stand securely during all switching processes.
Verify the changed damping characteristics after switching and observe the feedback from
the acoustic signal emitter.
Switch back to basic mode once the activities in MyMode have been completed.
Take the weight off the product and correct the switching, if required.
CAUTION
Improper use of the stance function
Falling due to unexpected behaviour of the product because of changed damping behaviour.
Make sure that you are standing safely when using the stance function and check the lock of
the knee joint before placing your full weight on the prosthesis.
Make sure you have been instructed in the correct use of the stance function by the O&P
professional and/or therapist. Information on the stance function see page 26.
16