Electromagnetic emissions
Interference meas­
urements
HF emissions accord­
ing to CISPR 11
Harmonics according
to IEC 61000-3-2
Voltage fluctu­
ations/flicker accord­
ing to IEC 61000-3-3
Electromagnetic interference immunity
Phenomenon
Electrostatic dis­
charge
High-frequency elec­
tromagnetic fields
Magnetic fields with
rated power frequen­
cies
Electrical fast transi­
ents/bursts
Surges
Line against line
Conducted interfer­
ence induced by high-
frequency fields
Voltage drops
Voltage interruptions
Compliance
Group 1/class B
Not applicable –
power below 75 W
Product meets the
requirements of the
standard.
EMC basic standard
or
test procedure
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-8
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
IEC 61000-4-11
IEC 61000-4-11
Electromagnetic environment directive
The product uses HF energy exclusively for its
internal functioning. Its HF emissions are
therefore very low, and interference with neigh­
bouring electronic devices is unlikely.
Interference immunity test level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air,
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
30 A/m
50 Hz or 60 Hz
± 2 kV
100 kHz repetition rate
± 0.5 kV, ± 1 kV
3 V
0.15 MHz to 80 MHz
6 V in ISM and amateur frequency bands
between 0.15 MHz and 80 MHz
80% AM at 1 kHz
0% U
; 1/2 period
T
At 0, 45, 90, 135, 180, 225, 270 and 315
degrees
0% U
; 1 period
T
and
70% U
; 25/30 periods
T
Single phase: at 0 degrees
0% U
; 250/300 periods
T