Successive
presses on
short
long (> 2 sec)
3.5.2 MAX/MIN Mode + Activation of the HOLD Mode
Successive
presses on
short
NOTE: The
MAX/MIN/PEAK values.
Clamp-On Meter Model 607- User Manual
- First press: Activates detection of
the MAX/MIN/AVG and PEAK values and
displays MAX/AVG and MIN values
- Second press: Displays the PEAK+, AVG,
and PEAK- values (on a second screen)
- Third press: Displays MAX/AVG and MIN
values without exiting from the mode (the
values already detected are not erased)
NOTE: Depending on the mode, AC or DC,
the crest factor (CF), harmonics, frequency,
and RIPPLE are also available.
- Activates the detection of MAX/MIN/ AVG
values
- Displays the MAX, MIN and AVG value
successively
- Returns to the display of the present
measurement without exiting from the mode
(the values already detected are not erased)
- Exits the MAX/MIN/PEAK mode. The
values previously recorded are then erased
NOTE: If the HOLD function is enabled, it is
not possible to exit from the MAX/MIN/ PEAK
mode. The HOLD function must first be
disabled.
- Displays the MAX, AVG, MIN and PEAK+,
AVG, PEAK- values detected before the
- When the
value is held on the display.
HOLD function does not interrupt the acquisition of new
Functions
Function
button was pressed.
button is pressed, the last
21