Dc Current Measurement; Ac Current Measurement - AEMC 607 User Manual

Clamp-on meter
Hide thumbs Also See for 607:
Table of Contents

Advertisement

Specific Specifications in MAX/MIN mode in Voltage
(from 10 Hz to 1 kHz in AC and AC+DC):
■ Accuracy: add 1 % R to the values of the previous table.
■ Capture of the extreme: approximately 100 ms.
Specific Specifications in PEAK mode in voltage
(from 10 Hz to 1 kHz in AC and AC+DC):
■ Accuracy: add 1.5 % R to the values in the previous table.
■ PEAK capture time: 1 ms min to 1.5 ms max.

5.2.4 DC Current Measurement

Measurement
Range
(2)
Specified
Measurement
Range
Accuracy
(2)
Resolution
(1) The display indicates +OL above 3000 A.
(2) The residual current at zero depends on the remanence. It can be corrected by the DC zero
function of the HOLD button.

5.2.5 AC Current Measurement

Measurement
Range
(2)
Specified
Measurement
Range
Accuracy
Resolution
(1) The display indicates OL above 3000 A (in PEAK mode).
The - and + signs are not displayed.
Bandwidth in AC = 1 kHz
(2) In AC, any value between zero and the min threshold of the measurement range (0.15 A) is
forced to show
Residual current at zero < 150 mA.
54
(0.00 to 99.99)
A
(0 to 100) % of the measurement range
± (1 % R + 10 cts) ± (1 % R + 3 cts)
0.01 A
(0.15 to 99.99) A
(0 to 100) % of the measurement range
± (1 % R + 10 cts)
0.01 A
on the display.
(100.0 to 999.9)
A
0.1 A
(100.0 to 999.9 A
± (1 % R + 3 cts)
0.1 A
Clamp-On Meter Model 607 - User Manual
(1000 to 3000)
A
(1)
2000 A
± (1.5 % R +3 cts)
(2000 to 2500) A
± (2.5 % R +3 cts)
(2500 to 3000) A
± (3.5 % R +3 cts)
1 A
(1000 to 2000) A
(1000 to 1500) A
± (1.5 % R +3 cts)
(1500 to 2000) A
± (2 % R +5 cts)
1 A
DC
DC
(1)

Advertisement

Table of Contents
loading

Table of Contents