Ac Voltage Measurement; Ac+Dc Voltage Measurement - AEMC 607 User Manual

Clamp-on meter
Hide thumbs Also See for 607:
Table of Contents

Advertisement

5.2.2 AC Voltage Measurement

Measurement Range
Specified
Measurement Range
Accuracy
Resolution
Input Impedance
(1) Above 1000 V (RMS), a repetitive beep indicates that the voltage being measured is greater
than the safety voltage for which the device is guaranteed. The display indicates OL.
Bandwidth in AC = 3 kHz.
(2) Any value between zero and the min threshold of the measurement range (0.15 V) is forced to
show
on the display.
Specific Specifications in MAX/MIN mode
(from 10 Hz to 1 kHz, and from 0.30 V in AC):
■ Accuracy: add 1 % R to the values of the table above.
■ Capture of the extreme: approximately 100 ms.

5.2.3 AC+DC Voltage Measurement

Measurement
Range
(2)
Specified
Measurement Range
Accuracy
Resolution
Input Impedance
(1) The display indicates OL above 1000 V (1400 V in PEAK mode).
Above 1000 V (DC or RMS), a repetitive beep indicates that the voltage being measured is
greater than the safety voltage for which the device is guaranteed.
Bandwidth in AC = 3 kHz.
(2) Any value between zero and the min threshold of the measurement range (0.15 V) is forced to
show
on the display.
Clamp-On Meter Model 607- User Manual
(0.15 to 99.99)
V
(0 to 100) % of the measurement range
(2)
(0.15 to 9.99) V
± (1 % R + 10 cts)
(10.00 to 99.99) V
± (1 % R + 3 cts)
0.01 V
(0.15 to 99.99)
V
(0 to 100) % of the measurement range
(0.15 to 9.99) V
± (1 % R + 10 cts)
(10 to 99.99) V
± (1 % R + 3 cts)
0.01 V
(100.0 to 999.9)
V
± (1 % R + 3 cts)
0.1 V
Ω
10 M
(100.0 to 999.9)
V
± (1 % R + 3 cts)
0.1 V
Ω
10 M
1000 V RMS
1400 V peak
(1)
1 V
1000 V RMS
(1)
1400 V peak
1 V
53