4.13.3 Individual Harmonics and Frequency of the
Fundamental from DC to the 25th
In the context of measurement of the THDs in voltage (§ 4.13.1) and in
current (§ 4.13.2):
1. Press the
only in DC or AC+DC. The harmonics of higher orders are displayed
sequentially as the
can be pressed to return to the previous order.
2. The
order of harmonic concerned.

4.14 RECORDING

The power meter allows recording of the data/measurements using the
REC function. The default recording interval is 60 secs. It can be configured
(see § 4.4.3) from 1 sec to 600 sec (10 mins).
1. First, select the function to be measured using the rotary switch,
then apply a long press (> 2 s) to the
displayed. Recording of the measurement starts. The data recorded is
in the format: MAX value – AVG Value – MIN Value – Unit – Mode
(AC or AC+DC).
46
button. Order hdC is displayed (DC component),
button is pressed repeatedly. The
button can be pressed to display the frequency of the
button. The REC symbol is
Clamp-On Meter Model 607 - User Manual
button