Peak; Min/Max Push Button: Recording Function - AEMC 721 User Manual

Clamp-on harmonic meter
Table of Contents

Advertisement

5.6.2 Peak

Peak is a fast responding function which measures the instantaneous
(1ms) Peak values of the measured signal. This function is often used for
half cycle in-rushing signals. The Peak function is displayed in the lower
right-hand corner of the display next to
the measurement symbol.
Peak may also be accessed when record-
ing. When recording MIN/MAX, depress-
ing PEAK gives the MAX Peak only.
The bargraph also indicates the Peak
values. When a Peak value exceeds
the bargraph scale, an arrow at the end
of the scale (►) is displayed. When in
MAX Peak, the MAX Peak value blinks
while the bargraph continues to respond
in Peak mode.
5.7

MIN/MAX Push Button: Recording Function

The recording feature allows the user to record the Minimum (MIN), Maxi-
mum (MAX), Average (AVG), and Maximum Peak (MAX Peak) measure-
ment values.
By pressing the MIN/MAX push button sequentially, the measurements
are displayed. Recording is possible in A, V, %THD, %DF, CF, and HZ.
If SMOOTH is selected prior to RECORDING, the smoothed values are
recorded.
Push Button Sequencing:
In A or V:
In A or V Peak:
In THD Functions: THD Function → Max → Min → Avg → THD Function
In Hz:
Note: The same operations apply with SMOOTH turned on
Recording is initiated with the MIN/MAX push button during the selected
measurement function. RECORD indication will blink 1/second during the
record mode. Control is verified with a short beep.
36
A or V Max → Min → Avg → A or V
A or V Peak → Max Peak → A or V Peak
Hz → Max → Min → Avg → Hz
���������
����
��������������
� � � � � � � � � � �
���
Harmonic Meter Model 721
����
����
���