Harmonic Measurement Specifications; Environmental Conditions - AEMC 607 User Manual

Clamp-on meter
Hide thumbs Also See for 607:
Table of Contents

Advertisement

5.2.24 Harmonic Measurement Specifications

Measurement Range in Voltage Per § 4.2.2 and § 4.2.3
Measure Range in Current
Harmonic Range
Frequency Analysis Band
Stability of the Current and
Voltage Display
Accuracy on the RMS value of
the harmonic (zero corrected
in A
)
DC
NOTE:
The display is
(V < 8 V or I < 9 A) or if the frequency is less than 5 Hz.
Specific characteristics in MAX/MIN mode in THD (from 10 Hz to 1 kHz):
■ Accuracy: add 1 % R to the values in the table above.
■ Capture time of the extreme: approximately 100 ms

5.3 ENVIRONMENTAL CONDITIONS

Conditions
Temperature
Relative Humidity (RH)
64
Per § 4.2.5 and § 4.2.6
AC: harmonics of orders 1 to 25
AC+DC: all orders from 1 to 25, plus the DC
component
- 0 to 25 times the fundamental frequency,
from among the network frequencies 50 Hz,
60 Hz, and 400 Hz
- 0 to 12 times the fundamental frequency of
an 800 Hz network
± (1 % R ± 2 cts)
Level >10 % and order <13: ± (5 % R ± 2 cts
Level >10 % and order >13: ± (10 % R ± 2 cts)
Level <10 % and order <13: ± (10 % R ± 2 cts)
Level <10 % and order >13: ± (15 % R ± 2 cts)
if the input signal is too low
Operating
(-4 to +131) °F
(-20 to +55) °C
≤ 90 % up to
131 °F (55 °C)
Clamp-On Meter Model 607 - User Manual
Storage
(-40 to +158) °F
(-40 to +70) °C
≤ 90 % up to
158 °F (70 °C)

Advertisement

Table of Contents
loading

Table of Contents