Calibration Procedure
Remarks
M1207026
M1207027
Reproduced from the electronic master in MATRIX
1. To start the Flow Sensor Positive Calibration, select "Start".
2. Reassemble the bellows assembly upon the completion of
calibration.
Ensure the flow sensor is correctly connected as detailed in the
calibration setup. The small port is the port closest to the patient
connection. The large port is the port closest to the patient circuit Y-
piece (Section 2.8.9).
The data is stored in EEPROM and is used during normal operation
to compensate for the individual flow sensor output characteristics.
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English
4 Service Tests and Service Mode
4-23