Tighten the mounting screws. Make sure that the top edge of
the switch assembly is parallel to the top edge of the gauge
panel.
10. Test the replacement switch assembly:
Connect the power cable to an electrical outlet.
Set the power switch to On.
Set the system switch to On.
Make sure that the display comes On.
Set the system switch to Standby.
Make sure that the display turns Off.
11. Reinstall the rear panel.
12. Perform the checkout procedure (Section 3).
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English