8.4 Software Installation

8-6
M1207027
Reproduced from the electronic master in MATRIX
1. Set the system switch to On.
2. Enter the Service Mode.
3. Select the "Software Upgrade" from the Configuration/
Upgrade menu.
4. Follow the instructions on the screen to complete the installation.
5. Set the system switch to Standby.
6. Remove the Software USB disk.
7. On the Function Upgrade menu (Section 4.6.1) of the Service
Mode, enter the options Key Code found on the side of the
display module to restore the original Upgrade Options.
8. On the Altitude Setting menu of the Service Mode, reset the
Altitude (Section 4.6.4).
9. Perform the preoperative checkout procedure.