4.4.4 Self - Test

Remarks
4-10
M1207027
Reproduced from the electronic master in MATRIX
The Self-Test menu allow you to observe the self tests and the
results on the screen.
The Self-Test shows:
CBD and DMB Communication — Pass or Fail
CSB and Alarm MCU Communication — Pass or Fail
Inspiratory Flow Valve (DAC) — Pass or Fail
Paw Sensor Reference voltage — Pass or Fail
Flow Sensor Reference voltage — Pass or Fail
DMB 3.3 V voltage — Pass or Fail
PEEP Valve (DAC) — Pass or Fail
Inspiratory Flow valve voltage — Pass or Fail
PEEP Valve (Drive) — Pass or Fail
10 VA of PEEP and Flow Valve — Pass or Fail
When the result is "Pass", the icon background should be green.
When the result is "Fail", the icon background should be red.
If one or more of these tests fail, please check the CSB, DMB, Vent
Engine and the harnesses.