9.12.3 Replace the
Bag port
M1207026
M1207027
Reproduced from the electronic master in MATRIX
1. Remove the breathing system.
2. Fully loosen the screw (A) to release the bag port.
3. Pull out the bag port from the manifold.
4. Replace the o-ring as necessary. Lubricate the o-ring sparingly
with Krytox.
5. Replace the bag port in reverse order.
6. Reinstall the breathing system.
7. Perform the checkout procedure (Section 3).
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English
9 Repair Procedure
A
O-ring
9-27