Discrete I/O Signal - GE 9100c Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

4.4.1 Discrete I/O
Signal
Mechanical switches
Pneumatic
Electronic switches
M1207026
M1207027
Reproduced from the electronic master in MATRIX
The Display Discrete I/O Signal menu displays discrete binary
Signal associated with machine switch positions Status.
There are several types of switches in the machine:
some switches are mechanically operated,
some switches are pneumatically operated,
some switches are electronic.
ACGO Switch — Vent or Aux CGO (machine equipped with
ACGO)
Canister Status — Closed or Open (Closed by default if
Canister Release switch not installed)
Bag/Vent Status — Vent or Bag
O
Supply Pressure — Pressure or No Pressure
2
O
flush — On or Off
2
O
Sensor Status — Connected or Disconnected
2
Breathing Circuit — Connected or Disconnected
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English
4 Service Tests and Service Mode
4-7