9100C Installation Checklist - GE 9100c Technical Reference Manual

Anesthesia machine
Table of Contents

Advertisement

9100c

6.1 9100c Installation Checklist

Serial Number:
Hospital:
6-2
M1207027
Reproduced from the electronic master in MATRIX
Date: (YY/MM/DD)
Performed by:
1. Unpack and assemble the 9100c System.
Note: install the AGSS first if equipped.
2. If necessary, access the Service Mode – Languages menu and
select the correct language.
3. Access the Service Mode – Configuration/Upgrade menu.
a. Enter the site Altitude in increments of 100 meters.
b. Enable – Disable installed optional monitoring and
ventilation features purchased by the customer (PCV mode,
ACGO switch, CO
4. Access the Service Mode – Calibration menu. Perform the
following calibrations:
a. set Calibration coefficient (Section 4.5.9).
b. Paw Zero (Section 4.5.1).
c. Paw Gain (Section 4.5.2).
d. O
Cell (Section 4.5.3).
2
e. PEEP Valve (Section 4.5.4).
f.
Flow Zero (Section 4.5.5).
g. Inspiratory Flow Valve (Section 4.5.6).
h. Flow Sensor Positive (Section 4.5.7).
i.
Flow Sensor Negative (Section 4.5.8).
5. Exit the Service Mode and go to Normal Mode.
6. Complete the System Checkout by performing the following
steps:
a. Inspect the system (Section 3.2).
b. Pipeline and cylinder tests (Section 3.3).
c. Flow control test (Section 3.4).
d. Pressure relief tests (Section 3.5).
e. O
supply alarm test (Section 3.6).
2
004
MANUAL−DOC, Technical Reference Manual, Technical Reference Manual for 9100c, English
bypass, O
monitoring, PEEP).
2
2
M1207026