Swann AMU-II Oxytrace Operator's Manual page 41

Table of Contents

Advertisement

AMU-II Oxytrace
Maintenance
Example
A-96.150.541 / 111223
As an example, the following picture shows the connection of the
AMI Inspector Oxygen via T-fitting to the process monitor.
A
AMI Inspector Oxygen
B
Flow cell of reference
instrument
C
Flow cell of online instrument
1 Stop sample flow to the online instrument by closing the appropri-
ate valve, e.g. back pressure regulator, sample preparation or
flow regulating valve at flow cell.
2 Connect sample line of the online instrument with the sample in-
let of the AMI Inspector Oxygen. Use the supplied tube made of
PA.
3 Connect sample outlet of the AMI Inspector Oxygen to the sam-
ple outlet funnel of the online instrument.
4 Switch on the AMI Inspector Oxygen. Open the flow regulating
valve and regulate the sample flow to 10 l/h. The flow rate is
shown on the display of the AMI Inspector Oxygen.
A
B
C
D
E
F
D
Sample outlet
E
Sample inlet
F
T-fitting
41