Clean Swansensor Oxytrace G And Flow Cell - Swann AMU-II Oxytrace Operator's Manual

Table of Contents

Advertisement

AMU-II Oxytrace
Maintenance
6.3.2
34

Clean Swansensor Oxytrace G and Flow Cell

Depending on the water quality, the Swansensor Oxytrace G and the
flow cell will necessitate a cleaning.
Before cleaning, proceed as described in
tenance, p.
31.
1 Dismount the Swansensor Oxytrace G, see
change, p.
32.
2 Clean the sensor with a soft tissue and rinse it with water after-
wards.
3 Use a soft brush to remove the dirt which sticks on the walls of
the flow cell.
4 Flush the flow cell with clean water.
5 Install the Swansensor Oxytrace G and start sample flow.
Stop of Operation for Main-
Electrolyte ex-
A-96.150.541 / 111223