AMU-II Oxytrace
Maintenance
Quality Level
0: Off
1: Trend
2: Standard
3: Crucial
4: User
a) sample temperature must be 25 °C +/-5 °C.
Procedure
6.6.1
A-96.150.541 / 111223
Limits and intervals:
max. deviation
temperature [°C]
off
0.5 °C
0.4 °C
0.3 °C
0–2 °C
The standard workflow consists of the following steps:
1 Activation of Swan quality assurance procedure
2 Pre-test
3 Connecting instruments
4 Carrying out comparison measurement
5 Completion of the measurement
Note: The procedure should only be carried out by qualified
personnel.
Activate Swan Quality Assurance Procedure
Activate the quality assurance procedure for the process monitor(s)
to be checked by selecting the quality level in menu 5.1.2.1. The cor-
responding submenus then become visible.
The activation is necessary the first time only.
max. deviation
a)
result [%]
off
10%
5%
5%
0–20%
min. inspection
interval
off
annual
quarterly
monthly
annual, quarterly,
monthly
39