Swann AMU-II Oxytrace Operator's Manual page 45

Table of Contents

Advertisement

AMU-II Oxytrace
Error List
Error
E001
Oxygen Alarm high
E002
Oxygen Alarm low
E003
Saturation Alarm high
E004
Saturation Alarm low
E007
Sample Temp. high
E008
Sample Temp. low
E009
Sample Flow high
E010
Sample Flow low
E011
Temp. shorted
E012
Temp. disconnected
A-96.150.541 / 111223
Description
Corrective action
– check process
– check programmed value,
see
5.3.1.1.1, p. 62
– check process
– check programmed value,
see
5.3.1.1.25, p. 62
– check process
– check programmed value,
see
5.3.1.4, p. 63
– check process
– check programmed value,
see
5.3.1.4, p. 63
– check process
– check programmed value,
see
5.3.1.3.1, p. 62
– check process
– check programmed value,
see
5.3.1.3.25, p. 62
– check sample flow
– check programmed value,
see
5.3.1.2.2, p. 62
– establish sample flow
– clean instrument
– check programmed value,
see
5.3.1.2.35, p. 62
– Check wiring of sensor
– Check sensor
– Check wiring of sensor
– Check sensor
45