Swann AMU-II Oxytrace Operator's Manual page 36

Table of Contents

Advertisement

AMU-II Oxytrace
Maintenance
36
Calibration
Place the electrode into
the wet flow cell at
a slightly tilted angle.
<Enter> to continue
Calibration
Saturation
98.7 %
Sat. Current
32 µA
Progress
Calibration
Saturation
98.7 %
Sat. Current
32 µA
Calibration Successful
7 Place the electrode slightly tilted into
3.1.5
the flow cell, so that the sensor cap
rests on the rim for the O-ring.
A Tilted sensor
B Flow cell
8 Press [Enter] to start the calibration
3.1.5
measurement.
 The saturation should reach
100%, the saturation current
should be about 22 µA to 33 µA.
If the measuring values are not
stable during the measuring
period, the calibration will be
discarded.
9 Press [Enter] to confirm the calibra-
3.1.5
tion.
A
B
A-96.150.541 / 111223