Swann AMU-II Oxytrace Operator's Manual page 69

Table of Contents

Advertisement

AMU-II Oxytrace
Program List and Explanations
5.4.2
5.4.3
5.4.4
5.4.5
5.5 Interface
5.5.1
5.5.20
5.5.30
5.5.40
5.5.1
5.5.21
5.5.31
5.5.41
5.5.1
5.5.1
A-96.150.541 / 111223
Set defaults: Reset the instrument to factory default values in three
different ways:
 Calibration: Sets calibration values back to default. All other
values are kept in memory.
 In parts: Communication parameters are kept in memory. All
other values are set back to default values.
 Completely: Sets back all values including communication pa-
rameters.
Load Firmware: Firmware updates should be done by instructed ser-
vice personnel only.
Password: Select a password different from 0000 to prevent unau-
thorized access to the menus "Messages", "Maintenance", "Opera-
tion" and "Installation".
Each menu may be protected by a different password.
If you forgot the passwords, contact the closest SWAN representa-
tive.
Sample ID: Identify the process value with any meaningful text, such
as KKS number.
Select one of the following communication protocols. Depending on
your selection, different parameters must be defined.
Protocol: Profibus
Device address:
ID-Nr.:
Local operation:
Protocol: Modbus RTU
Device address:
Baud Rate:
Parity:
Protocol: Hyperterminal
Baud Rate:
Protocol: HART
Address:
Range: 0–126
Range: Analyzer; Manufacturer; Multivariable
Range: Enabled, Disabled
Range: 0–126
Range: 1200–115 200 Baud
Range: none, even, odd
Range: 1200–115 200 Baud
0–63
69