Swann AMU-II Oxytrace Operator's Manual page 33

Table of Contents

Advertisement

AMU-II Oxytrace
Maintenance
A-96.150.541 / 111223
3 Unscrew and remove the sensor cap [H] from the Swansensor
Oxytrace G [A].
4 Empty the remaining electrolyte.
5 Refill the sensor cap with fresh electrolyte.
Note: There is a groove [C] in the thread [B] of the sensor, where
excessive air and electrolyte can escape while screwing the
sensor cap onto the sensor. Hold the sensor in vertical position,
measuring head pointing downwards.
6 Slowly screw the sensor cap onto the sensor to allow the excess
electrolyte to escape without building up too much pressure in-
side the electrode. Fasten sensor cap tightly.
7 Clean the sensor thoroughly and dry the sensing membrane with
a soft tissue.
8 Switch on power.
9 Let the sensor run-in at air for at least 30 min, better 1 h.
10 Afterwards perform an air calibration.
11 Install the sensor into the flow cell.
12 Select <Maintenance>/<Service>/<New Filling> to reset the
counter for remaining electrolyte.