Sensor Oxytrace G - Swann AMU-II Oxytrace Operator's Manual

Table of Contents

Advertisement

AMU-II Oxytrace
Product Description
2.2.2
Technical data
Temp. sensor
Measuring
range
Accuracy
Precision
Response time
Minimal flow
Pressure
resistance
Operating
temp.
Material
Protection
Weight
12

Sensor OXYTRACE G

Sensor for the measurement of dissolved oxygen in ultra pure water.
Precise oxygen measuring cell with integrated temperature sensor
and guard electrode for faster initial response time after maintenance.
Clark oxygen electrode
Cathode gold, anode silver, guard silver
Zero current-free electrode system
Robust 25 µm fluoropolymer diaphragm
NT5K
0–20 ppm O
(25 °C) or
2
0–200% saturation
0,3% if calibration temperature = measuring temperature
1,5% at ±10 °C deviation to cal. temperature
±1% of reading or ±0.15 ppb
t
< 30 seconds (rising concentration)
90
50 cm/s
3 bar
max. 50 °C
shaft: polyacetal copolymer
membrane: fluoropolymer
IP 68
150 g
59
100
A-96.150.541 / 111223