AMU-II Oxytrace
Product Description
Signal
outputs
Relays
Alarm relays
Input
Communica-
tion interface
(option)
Safety
features
10
Two signal outputs programmable for measured values (freely
scalable, linear, bilinear, log) or as continuous control output (control
parameters programmable).
Current loop:
Maximal burden:
Two potential-free contacts programmable as limit switches for
measured values, controllers or timer with automatic hold function.
Maximum load: 100 mA/50 V
Two potential-free contacts (one normally open and one normally
closed). Summary alarm indication for programmable alarm values
and instrument faults.
Normally open contact: closed during normal operation, open on
error and power loss.
Normally closed contact: open during normal operation, closed on
error and power loss
Maximum load: 100 mA / 50 V
One input for potential-free contact to freeze the measuring value or
to interrupt control in automated installations (hold function or remote
off).
RS485 interface (galvanically separated) for communication via
Modbus or Profibus DP
USB interface for logger download
HART interface
RS232 interface for logger download with HyperTerminal
No data loss after power failure. All data is saved in non-volatile
memory. Overvoltage protection of inputs and outputs. Galvanic sep-
aration of measuring inputs from signal outputs.
0/4–20 mA
510 Ω
A-96.150.541 / 111223