Carry Out Comparison Measurement - Swann AMU-II Oxytrace Operator's Manual

Table of Contents

Advertisement

AMU-II Oxytrace
Maintenance
6.6.4
42

Carry Out Comparison Measurement

1 Navigate to menu <Maintenance>/<Quality Assurance>.
2 Follow the dialog on the display.
Quality Assurance
- Carry out preparations
- Install Inspector
- Sample flow to 10 l/h
<Enter> to continue
Quality Assurance
Value O2
0.05 ppb
Value Temp.
25.00 C
Wait 10 Minutes
<Enter> to continue
Quality Assurance
Value O2
0.05 ppb
Value Temp.
25.00 C
Inspector O2
0.06 ppb
Inspector Temp.
25.0 C
<Enter> to continue
Quality Assurance
Value O2
0.05 ppb
Value Temp.
25.00 C
Inspector
0.06 ppm
Inspector Temp.
25.0 C
<Enter> to continue
Quality Assurance
Max. Dev. O2
Max. Dev. Temp.
Dev. O2
Dev. Temp.
QA-Check succesful
If the QA check is not successful, it is recommended to clean the
sensor. If the QA check fails again, contact your local Swan distribu-
tor for support.
3 Carry out pre test preparations
3.5.5
Connect instruments.
Regulate sample flow to 10 l/h us-
ing the appropriate valve.
4 Wait 10 minutes whilst measure-
3.5.5
ment is running.
Press [Enter] to continue.
5 Read the ppb value of the AMI In-
3.5.3
spector Oxygen and enter it in the
"Inspector O2" field.
Press [Enter] to confirm.
6 Read temperature value of the
3.5.4
AMI Inspector Oxygen and enter it
in the "Inspector Temp." field.
Press [Enter] to confirm.
Press [Enter] to continue.
 The results are saved in QA
3.5.5
history regardless if successful
0.5 %
or not.
0.4 °C
0.1 %
0.4 °C
A-96.150.541 / 111223